AEE 15 Watt Medical Series
15 Watt Medical DC-DC converter
The AEE15W-M series of 15 watt isolated DC-DC converters are designed for demanding instrumentation applications that require supplementary or reinforced insulation to comply with stringent industrial or medical safety standards.
These converters feature low leakage current and a reinforced insulation system – critical for medical equipment applications – rated for 300 Vac working voltage and 4,200 Vacrms input/output isolation. They comply with the UL/CSA/EN/IEC 60601-1 medical safety standard and are rated for 2X means of patient protection (MOPP).
AEE15W-M series single and dual output converters cover input voltages from 9 to 75 Vdc, with three 2:1 input voltage ranges of 9–18 V, 18–36 V and 36–75 V. Each of the three input voltage categories offers a choice of six models, which produce either single outputs of 5 V, 12 V, 15 V and 24V, or dual outputs of +/-12 V and +/-15V. Rated at 15 watts, the converters have an operating temperature range of -40 to 85˚C, with full power available up to 60˚C. Typical efficiency is 89% maximum.
The converters are fully encapsulated in an industry-standard 2 x1 inch (50.8 x 25.4 mm) low profile plastic case. Standard features include latching overvoltage and overtemperature protection, with hiccup mode overcurrent and short-circuit protection. All AEE15W-M series converters use 285 kHz fixed frequency switching and comply with EN 55022 class A and FCC level A standards for conducted and radiated EMI.
- Specifications
- Special Features
- Models - Stock check
-
Total Max Power:
15 W -
Total Max Current:
3 A -
Input Voltage:
9 V to 75 V -
Output Voltage:
-15 V to 24 V -
Outputs:
Single, Dual -
Size (L x W x H):
2.000" x 1.000" x 0.470" -
Warranty:
2 years -
Operating Temperature:
-40°C to 85°C
Electrical Specifications
- Input
- Output
- Isolation
- Other
- Input voltage range: 9V to 18V dc input
- Input voltage range: 18V to 36V dc Input
- Input voltage range: 36V to 75V dc Input
- Output voltage (Dual): +/-15V, +/-12V
- Output voltage (Single): 5V, 12V, 15V, 24V
- Isolation voltage: 4200 Vac isolation
- Insulation: Reinforced insulation
- Termination: Plated through hole termination
- Package: 2inch by 1inch
- Saftey approval: 60601 approved
- I/O isolation of 4200 Vac reinforced insulation rated for 300 Vac working voltage
- Medical safety to UL/CSA/EN/IEC60601-1 4th edition
- 2 *MOPP rated
- Wide 2:1 input voltage range
- Fully regulated output voltage
- No minimum load requirement
- Overload/voltage and short circuit protection
- Low leakage current <5 µA
- Operating temperature range –40 ˚C to +80 ˚C (with derating)
- Input filter meets EN55022, ClassA and FCC, Level A
- Medical EMC Standard meets 4th Edition of EMI EN55011 and EMS EN60601-1-2
- 2”x 1” plastic package
- 2 year product warranty