AEE 10W - Medical Series

10 Watt Medical Isolated DC-DC Converters

The AEE10W-M series of 10 watt isolated DC-DC converters are designed for demanding instrumentation applications that require supplementary or reinforced insulation to comply with stringent industrial or medical safety standards.

These converters feature low leakage current and a reinforced insulation system – critical for medical equipment applications – rated for 300 VAC working voltage and 4,200 VACrms input/output isolation. They comply with the UL/CSA/EN/IEC 60601-1 medical safety standard and are rated for 2X means of operator protection (MOOP).

AEE10W-M series single and dual output converters cover input voltages from 9 to 75 Vdc, with three 2:1 input voltage ranges of 9–18 V, 18–36 V and 36–75 V. Each of the three input voltage categories offers a choice of two models, which produce a fully regulated 5 V or 12 V output. Rated at 10 watts, the converters have an operating temperature range of -40 to 85˚C, with full power available up to 60˚C. Typical efficiency is 82%.

The converters are fully encapsulated in an industry-standard 2 x1 inch (50.8 x 25.4 mm) low profile aluminum case. Standard features include latching overvoltage and overtemperature protection, with hiccup mode overcurrent and short-circuit protection. All AEE10W-M series converters use 150 kHz fixed frequency switching and comply with EN 55022 class A and FCC level A standards for conducted and radiated EMI.

Datasheet Press Releases Technical Documentation Print This Page
  • Total Max Power:
    10 W
  • Total Max Current:
    1.6 A
  • Input Voltage:
    9 V to 75 V
  • Output Voltage:
    -15 V to 15 V
  • Outputs:
    Single, Dual
  • Size (L x W x H):
    2.000" x 1.000" x 0.400"
  • Warranty:
    3 years
  • Operating Temperature:
    -40 to +75 degC

Electrical Specifications

  • Input
  • Output
  • Logic Control
  • Isolation
  • Other
  • Input voltage: 9 - 18V
  • Input voltage: 18 - 36V
  • Input voltage: 36 - 75V
  • Output voltage: 5V
  • Output voltage: 12V
  • Enable: Positive enable
  • Isolation voltage: 4200Vac
  • Mechanical: Metal cased, encapsulated
  • Safety: IEC/EN 60601-1, 2 MOOP
  • - 4200VAC reinforced Insulation
  • - Insulation rated for 300VAC Working Voltage
  • - Medical Safety to UL/CSA/EN/IEC
  • - 60601-1 3rd Edition
  • - 2 MOOP rated
  • - Wide 2:1 Input Voltage Range
  • - Fully regulated Output Voltage
  • - Low Leakage Current
  • - Operating Temp. Range –40°C to 75 °C
  • - Input Filter meets EN 55022, class A and FCC, level A
  • - Overload Protection
  • - 2”x 1” Plastic Package
  • - 3 Years Product Warranty